Blood Bank Controller with Printer Output - Cooling Controller :

Overview / Features :

.

O) Data Logging Print Through Thermal Printer
O) Power Failure ( Battery Backup )
O) Low Battery Indication
O) Door Open

SPECIFICATION :

O) Data Logging Print Through Thermal Printer
O) Power Failure ( Battery Backup )
O) Low Battery Indication
O) Door Open
O) Microcontroller Based, Double display
O) Cooling Application
O) Input: RTD
O) Output: 1. Cooling Relay contact or SSR
                  2. HI-Low Alarm Relay contact or Buzzer
                 3. Defrost - Relay
O) Range: 00.0 to 10.0'C
O) Temperature setting lock as per range adjustable in set mode.
O) Alarm High 0.3 to 10.0 adjustable
O) Alarm Low 0.3 to 10.0 adjustable
O) Alarm mute by decrement key
O) Hysteresis 0.2 or 2 to 25.0c adjustable
O) Defrost Cycle start temperature adjustable
O) Defrost On time 0 to 250 Minute
O) Defrost Off time 0 to 9999 Minute
O) Defrost on indication by dp blinking
O) Defrost on time count down indication (Double display)
O) Selectable Print Time Off To 250 Minute.
O) Manual defrost by pressing increment key for 5 second
O) Delay 0 to 10 Minute adjustable
O) OPEn display if Sensor break
O) Pw.FL Display if Power Fail
O) Relay trip if sensor breaks or connection open.
O) Calibration Error Correction by front key switch ( No preset Setting )